White 18-liter Class B autoclave with USB and built-in printer
Warranty details
Safety
- It is essential that the water quality is below 20 microsiemens for the autoclave to start the cycles.
- The cycle will not start if the door is not properly closed.
- The automated safety system with a dual hydraulic locking mechanism prevents the door from opening while the chamber is under pressure.
- The safety valves prevent pressure generator and camera failure.
- A safety thermostat prevents overheating in the chamber and generator
- An automatic safety lock prevents the door from accidentally opening during the sterilization process. The process protector automatically cuts off the electrical power if there is a short circuit anywhere or if the power supply is unstable.
- The failure alarm indicates a failed cycle and the detailed error code.
- The door alarm indicates that the door is not secured.
- The low water alarm indicates when the water tank is empty or overflowing.
Technical specifications
- Voltage: 220-240 VAC; 50/60 Hz; 10A
- External dimensions of the device: W x H x L: 490 x 455 x 600 mm
- Tray dimensions: 285 x 192 mm
- Circuit breaker: F16A / 400 V
- Rated power: 1500 VA
- 1750 W
- Sterilization temperature: 121 °C / 134 °C
- Distilled water tank capacity: 2.5 L
- Operating temperature: 5 °C / 40
- Relative humidity: Max. 80%, non-condensing
- Net weight: 47 kg
- Maximum noise level: <70 dB
- Atmospheric pressure: 76 kPa / 106 kPa
Product intended for the European market – 220–240 V / 50 Hz power supply – European plug (type E/F)
Icanclave Autoclave: Maintenance and Error Message Guide
Components
- Distilled water tank
- LCD screen
- Control panel
- Main switch
- Hose for draining distilled and used water
- Key for tightening the door
- USB port
- Internal printer (allows printing of sterilization cycle reports on thermal paper after the end of the cycle.)
- Safety valve
- Ventilation grille
- Locking system
- Power cable
- Label
Included accessories
- Sterilizer
- USB
- 4 instrument trays
- Tray support
- Handle for removing trays
- Door adjustment tool
- 2 drain hoses
- Door key
- Instruction manual
Regulations
- 93/42 / EEC Medical devices directive.
- 97/23 / EEC Pressure equipment directive.
- EN 13060 Small steam sterilizers.
- EN 61326, EN 61000 EMC.
- EN 61010-1, UL 61010-1 safety requirements.
- EN 61010-2-040, IEC 61010-2-040 Particular requirements for steam sterilizers used for processing medical materials.
Autoclave process step by step
1. Open the door, take the tools and accessories located in the sterilization chamber, remove the packaging and clean them.
2. Connect the power cord to a suitable power source.
3. Connect the printer.
4. Power on: the main switch is located on the front part, at the bottom right of the device. Once turned on, the LCD screen will light up and the basic operating symbols of the sterilizer, such as the door position, water level, sterilization program, date, time will light up.
5. Remove the top cover then fill the distilled water tank. If an audible signal sounds, it means you have exceeded the allowed water level in the tank. Empty the excess water until the alarm stops.
6. Preparation of the material to be sterilized: place the instruments on the different trays, and separate them according to their materials. If the instruments are not made of stainless steel, use sterilization pouches. These must not be in contact. Check that all instruments are properly placed, so that water does not remain in hollow parts. Do not overload the instrument trays and do not place them in direct contact with the walls of the sterilization chamber. Do not use paper clips or needles to avoid any damage or malfunction of the autoclave.
7. Select the sterilization program.
8. Start the sterilization cycle.
After selecting the sterilization program, place the instruments on the trays in the chamber using the provided plastic tongs.
Once the instruments are properly placed inside, close the door. Caution: make sure to close the door properly, otherwise the sterilizer will not be able to operate.
9. Start the sterilization program.
Press the start button, the autoclave will begin to operate for about 30-75 minutes.
10. End of the sterilization cycle.
Once the sterilization cycle is complete, the printer will automatically print the full cycle report.
Once the pressure has dropped back to 0, you can open the door and remove the instruments. The autoclave will alert you with an audible signal once the sterilization cycle is complete.
Caution: Use the plastic tongs to remove the instruments.
IMPORTANT: Do not reuse the wastewater after draining the tank.
It is important to keep the packaging box in case of after-sales service
| Program |
Temp.
/Pressure
|
Sterilization time |
Total time (min) |
| 8D Pro |
12D Pro |
18D Pro
|
23D Pro |
| QUICK 134 |
134℃/210Kpa |
3.5 min |
10-20 |
11-20 |
12-30 |
15-30 |
| PACKAGED 134 |
134℃/210Kpa |
4 min |
25-40 |
28-45 |
30-49 |
38-57 |
| PACKAGED 121 |
121℃/110Kpa |
20 min |
36-46 |
37-48 |
38-54 |
46-62 |
| PRION 134 |
134℃/210Kpa |
18 min |
35-46 |
36-48 |
37-52 |
45-60 |
| TEST B&D |
134 |
3.5 min |
10-20 |
15-25 |
15-30 |
28-35 |
| TEST HELIX |
134 |
3.5 min |
10-20 |
15-25 |
15-30 |
28-35 |
| EMPTY TEST |
– |
– |
17-20 |
18-21 |
19-22 |
20-25 |
* The cycle duration is based on different loads and with/without preheating.
* The sterilization time and drying time can be adjusted.